FLUKE 1587FC
- Stock: In Stock
- Model: Insulation Multimeter (50V, 100V, 250V, 500V, 1000V)
Overview:
The Fluke 1587 FC Insulation Multimeters combine a digital
insulation tester with a full-featured, true-rms digital multimeter in a single
compact, handheld unit, which provides maximum versatility for both
troubleshooting and preventative maintenance.
The Fluke 1587 FC Insulation Multimeter adds four powerful
new diagnostic functions through the Fluke Connect® Measurements app:- PI/DAR
timed ratio tests with TrendIt™ graphs identifies moisture and
contaminated insulation problems faster
- Memory
storage through Fluke Connect eliminates writing down results, reduces
errors and saves data for historical tracking over time
- Temperature
Compensation through app for establishing accurate baselines and relevant
historical comparisons
- Historical
tracking and trending of assets identifies degradation over time, allows
real-time decisions to be made in the field with Fluke Connect® Assets
Product highlights- PI/DAR
timed ratio tests
- Live
circuit detection prevents insulation test if voltage > 30 V is detected
for added user protection
- VFD
low-pass filter for accurate motor drive measurements
- Auto-discharge
of capacitive voltage for added user protection
- Insulation
test (1587 FC: 0.01 MΩ to 2 GΩ)
- Insulation
test voltages (1587 FC: 50 V, 100 V, 250 V, 500 V, 1000 V) for many
applications
- AC/DC
voltage, DC millivolts, AC/DC milliamps, Resistance (Ω), Continuity
- Capacitance,
diode test, temperature, Min/Max, frequency (Hz)
- Auto
power off to save battery power
- CAT
III 1000 V, CAT IV 600 V measurement category
- Large
display with backlight
- Rugged,
utility hard case allows you to bring everything you need for the job
- Included
accessories: Remote probe, test leads and probes, alligator clips, K-type
thermocouple
- Accepts
optional Fluke TPAK Magnetic Meter Hanger for convenient hands free
operation
Features:
- PI/DAR
measurements with TrendIt graphs
- Memory
storage through Fluke Connect Measurements app
- Temperature
Compensation through Fluke Connect Measurements app
- VFD
low-pass filter for accurate motor drive measurements
- Insulation
test voltages 50 V, 100 V, 250 V, 500 V, 1000 V
- Insulation
test: 0.01 MΩ to 2.0 GΩ
- Auto-discharge
of capacitive voltage
- Insulation
test smoothing reading
- Frequency
- Capacitance
- Diode
test
- Temperature
- Min/Max
- AC/DC
Voltage
- DC
Millivolts
- AC/DC
Milliamps
- Resistance
(0.1 Ω to 50 MΩ)
- Continuity
- Remote
probe, test leads, alligator clips
- K-type
thermocouple
- Rugged,
utility hard case
- Auto
power off
Specifications:
Electrical Specifications |
||
AC voltage measurement |
||
Range |
||
600.0 mV |
Resolution |
0.1 mV |
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) |
±(1% + 3) |
|
Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) |
±(2% + 3) |
|
6.000 V |
Resolution |
0.001 V |
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) |
±(1% + 3) |
|
Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) |
±(2% + 3) |
|
60.00 V |
Resolution |
0.01 V |
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) |
±(1% + 3) |
|
Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) |
±(2% + 3) |
|
600.0 V |
Resolution |
0.1 V |
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) |
±(1% + 3) |
|
Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) |
±(2% + 3)¹ |
|
1000 V |
Resolution |
1 V |
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) |
±(2% + 3) |
|
Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) |
±(2% + 3)¹ |
|
¹ 1 kHz bandwidth |
||
Low-Pass Filter Voltage |
||
Range |
||
600.0 mV |
Resolution |
0.1 mV |
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) |
±(1% + 3) |
|
Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) |
+(2% + 3), -(6% - 3) |
|
6.000 V |
Resolution |
0.001 V |
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) |
±(1% + 3) |
|
Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) |
+(2% + 3), -(6% - 3) |
|
60.00 V |
Resolution |
0.01 V |
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) |
±(1% + 3) |
|
Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) |
+(2% + 3), -(6% - 3) |
|
600.0 V |
Resolution |
0.1 V |
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) |
±(1% + 3) |
|
Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) |
+(2% + 3), -(6% - 3) |
|
1000 V |
Resolution |
1 V |
Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) |
±(2% + 3) |
|
Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) |
+(2% + 3), -(6% - 3) |
|
DC Voltage Measurement |
||
Range |
Resolution |
Accuracy ±(% of Rdg + Counts) |
6.000 V dc |
0.001 V |
±(0.09% + 2) |
60.00 V dc |
0.01 V |
±(0.09% + 2) |
600.0 V dc |
0.1 V |
±(0.09% + 2) |
1000 V dc |
1 V |
±(0.09% + 2) |
Input impedance |
10 MΩ (nominal), <100 pF |
|
Normal mode rejection ratio |
>60 dB @ 50 Hz or 60 Hz |
|
Common mode rejection ratio |
>120 dB @ dc, 50 Hz or 60 Hz (1 k unbalance) |
|
Accuracies apply to ± 100% of range |
||
DC Millivolts Measurement |
||
Range |
Resolution |
Accuracy ±(% of Rdg + Counts) |
600.0 mVdc |
0.1 mV |
±(0.1% + 1) |
DC and AC Current Measurement |
||
AC 45 Hz to 1000 Hz |
||
Range |
400 mA |
|
Resolution |
0.1 mA |
|
Accuracy ±(% of Rdg+Counts) |
±(1.5% + 2)¹ |
|
Burden voltage (Typical) |
2 mV/mA |
|
Range |
60 mA |
|
Resolution |
0.01 mA |
|
Accuracy ±(% of Rdg+Counts) |
±(1.5% + 2)¹ |
|
Burden voltage (Typical) |
2 mV/mA |
|
DC |
||
Range |
400 mA |
|
Resolution |
0.1 mA |
|
Accuracy ±(% of Rdg+Counts) |
±(0.2% + 2) |
|
Burden voltage (Typical) |
2 mV/mA |
|
Range |
60 mA |
|
Resolution |
0.01 mA |
|
Accuracy ±(% of Rdg+Counts) |
±(0.2% + 2) |
|
Burden voltage (Typical) |
2 mV/mA |
|
Overload |
600 mA for 2 minutes maximum |
|
Fuse protection for mA input |
0.44 mA, 1000 V, IR 10 kA |
|
AC conversion |
Inputs are ac-coupled and calibrated to the rms value of sine wave input |
|
Conversions are true-rms responding and specified from 5% to 100% of range. Input signal crest factor can be up to 3 up to 300 mA, decreasing linearly to crest factor ≤1.5 at 600 mA. For non-sinusoidal waveforms add +(2% reading + 2% FS) typical, for a crest factor up to 3. |
||
¹ 1 kHz bandwidth |
||
Ohms Measurement |
||
Range |
Resolution |
Accuracy +(% of Rdg+Counts)¹ |
600.0 Ω |
0.1 Ω |
±(0.9% + 2) |