DSO 5202B HANTEK

DSO 5202B HANTEK
  • Stock: Pre-Order
  • Model: Digital Storage Oscilloscope - 200MHz
75,000
No Shipping Charges
ModelDSO5202BDSO5102BDSO5062B
Acquisition
Sample RateReal-Time Sample: 1GS/s
Equivalent Sample: 25GS/s
Acquisition Modes
NormalNormal data only
Peak DetectHigh-frequency and randon glith capture
AverageWavefom Average, selectable 4,8,16,32,64,128
Inputs
Inputs CouplingAC, DC, GND
Inpits Impendance1MΩ±2% 20pF±3pF
Probe Attenuation1X, 10X
Supported Probe Attenuation Factor1X, 10X, 100X, 1000X
Maximum Input VoltageCAT I and CAT II: 300VRMS (10×), Installation Category;
CAT III: 150VRMS (1×);
Installation Category II: derate at 20dB/decade above 100kHz to 13V peak AC at 3MHz* and above. For non-sinusoidal waveforms, peak value must be less than 450V. Excursion above 300V should be of less than 100ms duration. RMS signal level including all DC components removed through AC coupling must be limited to 300V. If these values are exceeded, damage to the oscilloscope may occur.
Horizontal
Sample Rate Range500MS/s--1GS/s
Waveform Interpolation(sin x)/x
Record LengthMaximum 1M samples per single-channel;
maximum 512K  samples per dual-channel (4K,40K optional)
SEC/DIV Range2ns/div to 40s/div
Sample Rate and
Delay Time Accuracy
±50ppmat over any ≥1ms time interval)
Position Range2ns/div to 8ns/div;
(-8div x s/div) to 20ms;
20ns/div to 80us/div; (-8div x s/div) to 40ms;
200us/div to 40s/div; (-8div x s/div) to 400s
Delta Time Measurement Accuracy
(Full Bandwidth)
Single-shot, Normal mode:± (1 sample interval +100ppm × reading + 0.6ns);
>16 averages:± (1 sample interval + 100ppm × reading + 0.4ns);
Sample interval = s/div ÷ 200
Vertical
Vertical Resolution8-bit resolution, all channel sampled simultaneously
Offset Range2mV/div to 20mV/div, ±400mV;  50mV/div to 200mV/div, ±2V
500mV/div to 2V/div, ±40V; 5V/div to 10V/div, ±50V
Bandwidth200MHz100MHz60MHz
Rise Time at BNC( typical)1.8ns 3.5ns5.8ns
Math+, -, *, /, FFT
FFTWindows: Hanning, Flatop, Rectamgular, Bartlett, Blackman;
1024 sample point
Bandwidth Limit20MHz
Low Frequency Response (-3db)≤10Hz at BNC
DC Gain Accuracy±3% for Normal or Average acquisition mode, 10V/div to 10mV/div;
±4% for Normal or Average acquisition mode, 5mV/div to 2mV/div
DC Measurement Accuracy,
Average Acquisition Mode
When vertical displacement is zero, and N ≥16:± (3% × reading + 0.1div + 1mV) only 10mV/div or greater is selected;
When vertical displacement is not zero, and N≥16: ± [3% × (reading + vertical position) + 1% of vertical position + 0.2div]; Add 2mV for settings from 2mV/div to 200mV/div; add 50mV for settings from 200mV/div to 10V/div
Volts Measurement Repeatability,
Average Acquisition Mode
Delta volts between any two averages of ≥16 waveforms acquired under same setup and ambient conditions