ZF ZHONGFENG
- Stock: In Stock
- Model: Electrolytic Capacitor Withstand Voltage Tester 200V/600V
Voltage device to stand electrolytic voltage
0-600v, current: 2-100ma, charged and recharged
Overview
Electrolytic
capacitance tester instructions:
Technical parameters:
Input voltage: ac 220v
Test voltage: 200v 600v
Test current: 2, 5, 50, 100ma
Instructions:
1. Turn on the power switch and put k3
in the discharge position.
2. If the standing voltage of the
capacitor to be tested is within 200v, place k2 in the 200V range, if it is
greater than 200V, place k2 in the 600V range, and place k1 in the appropriate
test current range.
3. Insert the capacitor under test into
the Cx socket. The positive and negative dice correspond to + and - of CX.
4. Turn on switch k3 to the pressure
measurement position, read the position voltage value after the displayed value
stabilizes, then turn k3 to the discharge position, and then remove the
capacitor.
5. Note that the capacitor should be
discharged before testing. If the high-voltage capacitance exceeds 4700
microfarads, please do not directly discharge it on the machine to avoid damage
to the machine.
1:
chip capacitors or ceramic capacitors
2: The current setting mainly determines the test speed. The larger the
capacity, the higher the current.