Main Menu / Categories
Your Cart

FLUKE 750R31

Buy Now Fluke 750R31 Reference Pressure Module At Best Price In Pakistan 2025

The Fluke 750R31 Reference Pressure Module is designed for high-pressure calibration applications requiring precise, stable, and traceable performance. With a wide range of 0 to 10,000 psi (0 to 700 bar / 70 MPa), it is suitable for demanding industrial environments including oil and gas, power generation, aerospace, petrochemical, and heavy manufacturing.

This module features internal temperature compensation for full-accuracy operation over a broad temperature range, ensuring reliable measurements in field and plant conditions across Pakistan. Constructed with 316 stainless steel for media compatibility and fitted with a rugged urethane-molded protective case, the 750R31 is built for long-term durability.

Fully compatible with Fluke 740 Series, 750 Series, and select 720/710 Series calibrators, the module offers reference-class performance with a total uncertainty of 0.015 percent. Each module includes a NIST-traceable calibration certificate and carries its own serial number for independent traceability.


Key Features

  • Pressure range from 0 to 10,000 psi (0 to 700 bar / 70 MPa)

  • Compatible media: 316 stainless steel

  • Displays pressure in any of 10 selectable pressure units through the calibrator

  • Rugged urethane molded case for protection in harsh environments

  • Internal temperature compensation from 32°F to 122°F for accurate field performance

  • NIST-traceable calibration certificate included

  • Can be used with Fluke 740, 750, and compatible 720 or 710 Series calibrators

  • Reference-class total uncertainty of 0.015 percent

  • Each module has its own serial number for independent traceability

  • Includes adapter set for flexible connection options


Specifications

  • Parameter Range: 0 to 10,000 psi (0 to 700 bar)

  • Burst Rating: 2 times full scale

  • High Side Media Compatibility: 316 stainless steel

  • Low Side Media: Not applicable

  • Reference Uncertainty: ±0.01 percent of full scale

  • Total Uncertainty (1-year, 15–35°C): ±0.02 percent of full scale

  • Total Uncertainty (1-year, full temperature range): ±0.04 percent of full scale

  • Total Uncertainty (6-month, 15–35°C): ±0.015 percent of full scale

  • Total Uncertainty (6-month, full temperature range): ±0.035 percent of full scale