HT824 Hantek
- Stock: Pre-Order
- Model: Process Calibrator
HT-824 multi functional process calibrator has high accuracy, high
stability, can be widely used in site calibration of industry instruments and
equipment in laboratory. Powerful functions, it can simulate to output voltage
in milli volt and volt, current in milliampere, and many types of electric
signals needed by the measurement and control during industry control process.
It can also test or simulate kinds of TC and RTD signals.
·
HT824 multi functional process
calibrator has high accuracy, high stability, can be widely used in site
calibration of industry instruments and equipment in laboratory. Powerful
functions, it can simulate to output voltage in millivolt and volt, current in
milliampere, and many types of electric signals needed by the measurement and
control during industry control process. It can also test or simulate kinds of
TC and RTD signals.
·
5 1/2 digit input/output
display.
·
Manual/auto cold junction
compensation and setting.
·
The temperature value can be directly
measured/output.
·
Auto power off when the voltage
of battery is low.
·
Apply advanced built-in battery
management system. 7.4V/7800mAh lithium battery, which can easily last for
over 8 hours during site measurement work.
·
Auto triangular wave/step wave
signal output.
·
It can store 8 sets of
measurement and output data, and recall it to display. All the setting value
and the function in the last operation can be saved automatically after
power-off.
·
The screen shows the connection
diagram of probes and plugs to avoid wrong operation. It has backlight.
·
The output has manual/auto step
and slope output function. It can arbitrarily set the value of 0%, 25%, 100% in
the range of signal output to test the response speed of transmitter.
·
Application field: measurement
and calibration of instruments, maintenance and overhaul of temperature
instruments, automatically maintenance and overhaul, project construction
debugging, detection and diagnosis of control system signal source,
configuration of site instrument repairmen room, matching of the thermal
engineering laboratory in electric power industry.
Model |
HT824 |
||||||
DC Voltage Measurement |
|||||||
Range |
0-30.000V(Upper part of screen)*1 |
0-24.000V(Bottom of screen)*2 |
0-90.000mV |
||||
Resolution |
0.001V |
0.001V |
0.001mV |
||||
Accuracy |
0.1%+5 |
0.05%+5 |
0.05%+5 |
||||
*1和*2:Input resistance is greater than 1MΩ |
|||||||
DC Voltage Output |
|||||||
Range |
0—20.000V |
0—90.000mV |
|||||
Resolution |
0.001V |
0.001mV |
|||||
Accuracy |
0.05%+5 |
0.05%+5 |
|||||
Millivolt measurement and output*1 |
|||||||
Range |
-10.000mV—80.000mV |
||||||
Resolution |
0.001mV |
||||||
Accuracy |
0.05%+5 |
||||||
*1 Press TC(16) to select this function. Signal is at thermocouple micro input/output TC port. |
|||||||
DC Current (milliampere) Measurement |
|||||||
Range |
0—24.000mA(Upper part of screen)*1 |
0—24.000mA(Bottom of screen)*2 |
|||||
Resolution |
0.001mA |
0.001mA |
|||||
Accuracy |
0.05%+5 |
0.05%+5 |
|||||
DC Current (milliampere) Output |
|||||||
Range |
0—24.000mA |
||||||
Resolution |
0.001mA |
||||||
Accuracy |
0.05%+5 |
||||||
SIMU(Simulation)When outputting current, external voltage is greater than 12V, less than 28V. |
|||||||
SOUR(Source)When outputting current, signal driving ability is 1000Ω in 20mA. |
|||||||
Resistance Measurement |
|||||||
Ohm Range |
|
0—400.00Ω |
400.0—4000.0Ω |
||||
Accuracy ±Ω |
4 Wire(4W) |
0.02%+5 |
0.02%+5 |
||||
2 Wire(2W)and 3 Wire(3W) |
0.05%+5 |
0.05%+5 |
|||||
Excitation current:0.2mA。 |
|||||||
Maximum input voltage:30V。 |
|||||||
2 Wire:Do not include wire resistance. |
|||||||
3 Wire:Assume to use the matched testing wire, the total resistance cannot be greater than 100Ω. |
|||||||
Resistance output |
|||||||
Range |
5.00Ω―400.00Ω |
400.0―1500.0Ω |
|||||
Excitation current from the measurement instrument |
0.15mA-2mA |
0.05mA-0.8mA |
|||||
Accuracy ±Ω |
0.05%+5 |
0.05%+5 |
|||||
Resolution |
0.1Ω |
0.1Ω |
|||||
Temperature - Thermocouple |
|||||||
Type |
Range |
Resolution |
Measurement and output accuracy ±℃ |
||||
S |
-50.0℃ to 0℃ |
0.1℃/0.1℉ |
2℃ |
||||
0℃ to 500.0℃ |
0.1℃/0.1℉ |
1.5℃ |
|||||
500.0℃ to 1760.0℃ |
0.1℃/0.1℉ |
1.3℃ |
|||||
R |
-50.0℃ to 0℃ |
0.1℃/0.1℉ |
2℃ |
||||
0℃ to 500.0℃ |
0.1℃/0.1℉ |
1.5℃ |
|||||
500.0℃ to 1760.0℃ |
0.1℃/0.1℉ |
1.3℃ |
|||||