ES8020 FUZRR






- Stock: Pre-Order
- Model: Battery Internal Resistance Tester
Temperature resolution: 0.1℃
Response time:200ms
Measure time: About 2 seconds
USB interface: With a USB interface, the stored data can be
uploaded to the computer, saved and printed
Bluetooth connection:Yes
Function | Battery internal resistance measurement, battery voltage
measurement, temperature measurement |
Accuracy guaranteed temperature and humidity | 23℃±5℃,
Below 75%rh |
Power supply | DC 3.7V lithium battery |
Resistance resolution | 1 u Ω |
Voltage resolution | 1 mV |
Temperature resolution | 0.1℃ |
Measuring range | Internal resistance measurement:0.000mΩ~3.100 Ω (4
ranges configuration) |
| Voltage measurement: 0.000V~±120.00V ( 2range
configuration) |
| Temperature measurement: -10.0℃~60.0℃
(single range configuration) |
Maximum input voltage | DC 70V (between + measurement terminal
and – measurement terminal), no AC input |
Measurement method | Internal resistance measurement: 1KHz AC 4-terminal test
method, open-circuit terminal voltage 3V max Measuring current: 2.0mA~200mA (different measuring
currents in different ranges) Temperature measurement: NTC temperature sensor (10KΩ at
26°C) A/D conversion method: successive approximation type Display update frequency: 5 times/second |
Response time | 200ms |
Measure time | About 2 seconds |
LCD Size | 70.1mm×52.6mm /3.5 inches (320*240 resolution 16-bit true
color screen) |
Instrument size | 190mm×121mm×51mm |
USB interface | With a USB interface, the stored data can be uploaded to
the computer, saved and printed |
Bluetooth connection | Yes |
Hold and store function | Manual retention and storage, automatic retention and
storage |
Measurement judgment function | Pre-set PASS, WARNING, FAIL judgment thresholds |
Battery voltage | The battery level is displayed in 5 bars, reminding to
charge in time when the battery voltage is low |
Automatic shut-down | No operation when power on, it will power off
automatically after about 15 minutes (can be turned off in the settings) |
Power consumption | 300mA MIN / 500mA MAX |
Weight | Meter:480g (With battery) |
Working temperature and humidity | -10℃~40℃;below 80%RH |
Storage temperature and humidity | -20℃~60℃;below 70%RH |
Insulation resistance | Above 20MΩ (500V between circuit and case) |
Withstand voltage | AC 3700V/RMS (between circuit and case) |
External magnetic field | <40A/m |
External electric field | <1V/m |
Suitable for safety regulations | IEC 61010 |